This chapter answers one question: what is the camera actually measuring, and what makes the number wrong?

In Fahrenheit: with the SDK switched to Fahrenheit (SetTemperatureUnit() and its counterparts, see Temperatures in Fahrenheit), every temperature on this page is read and written in Fahrenheit, and every difference in Fahrenheit degrees; the samples are unchanged.

It assumes no thermography. Nothing in it is specific to this SDK; it is the physics and the operation of an infrared camera, and it is here because every chapter after it uses the vocabulary. If you have measured with thermal cameras before, read the glossary at the end and move on.

A thermal camera does not measure temperature

It measures radiation. Every surface above absolute zero radiates infrared energy; the camera's detector collects that energy over a short exposure and reports, for each of its elements, a number proportional to how much arrived. Turning that number into a temperature is an inference, and radiometry is the name for doing it properly.

The chain inside the camera and the SDK runs in four stages, and every chapter in this manual that talks about pixels is talking about one of them:

Stage What a pixel holds Where you see it
Raw The detector's own uncalibrated reading, in ADU: analogue-to-digital units, the integer that came out of the converter. Not a temperature, not even proportional to one RawFrame
Energy A linear measure of radiant energy, after the per-pixel corrections and the radiation parameters have been applied. Still not a temperature EnergyFrame
Temperature Energy looked up through the camera's own characteristic curve: the calibration measured for that unit, that optic and that temperature range ThermalFrame
Picture The temperature mapped through a colour palette, for a human BgraFrame

Two consequences are worth carrying through the rest of the manual.

The calibration is per unit, not per model. The characteristic curve that turns energy into a temperature was measured on your camera with that optic in that range. It is why the SDK needs calibration files, why it will go and fetch them, and why a camera whose files cannot be found still opens and streams a perfectly good picture while its temperatures are wrong. Calibration File Acquisition is the chapter on that.

Raw and energy frames are in sensor coordinates. They are never rotated or mirrored, because the per-pixel calibration data they would be compared against is in sensor coordinates too. Temperature and picture frames are oriented as you asked.

The three settings that decide the number

The camera can only measure the energy that arrives at it. How much of that energy came from the object (rather than from the object's surroundings reflecting off it, or from the air and the window in between) is something you have to tell the camera. There are three settings, they are set together as one record on Radiometry(), and they are the single largest source of a wrong reading.

Emissivity (ε)

How efficiently a surface radiates, from 0 to 1. A perfect radiator, a "black body", is 1.0. Real surfaces are less, and a polished metal surface is nowhere near it. The SDK's default is 1.0, which assumes a perfect radiator; a reading taken at the default on anything else is wrong, and wrong in one direction, low.

Emissivity is not a cosmetic knob. On a real PI 640i, measured entry for entry through that camera's own calibration, a scene the camera reads as 50.00 °C at emissivity 0.95 reads 51.20 at 0.90, 54.30 at 0.80 and 70.00 at 0.50. Twenty kelvin, from one number, on the same unchanged scene.

This is also why emissivity cannot be a global application setting if you are measuring more than one material. A region can carry its own emissivity, and Radiometry().TemperatureAtEmissivity() re-evaluates a reading you already have at a different value without disturbing the camera. Frame Processing Settings covers both, and Regions covers per-region values.

Transmissivity (τ)

How much of the object's radiation survives the path to the camera, again 0 to 1. 1.0 is a clear path and is the default. Anything you are looking through (a window, a protective pane, a sight glass) absorbs some of the signal and radiates some of its own, and a camera told nothing about it reads the object as colder than it is. Ordinary window glass is close to opaque in the infrared, which is why thermal cameras look through germanium or special IR panes and not through glazing.

Ambient temperature, which is also the reflected temperature

The temperature of the surroundings, in °C, default 20.0. It does two jobs at once, and the SDK deliberately holds one value for both rather than two that would have to agree:

  • the temperature of the air in the path, which radiates and absorbs along with the window above, and
  • the reflected reference: a low-emissivity surface is partly a mirror, so some of what the camera sees is the room. Correcting for that needs to know how warm the room is.

So there is one ambient_temp_c and no separate reflected-temperature setting. There are also no distance or humidity settings; the SDK's model does not take them.

The camera can work the ambient out for itself from its housing probe, with estimate_ambient. Do that rather than guessing, if you have nothing better. Two things follow: reading ambient_temp_c back always gives what you last wrote, never what the camera decided, so GetEffectiveAmbientTemperature() is the one to read for the value actually in force; and the estimate steps rather than drifts, because it carries a 0.1 K hysteresis, so it cannot be reproduced from outside the SDK and should be read rather than recomputed.

The camera's own four temperatures

A camera reports several temperatures about itself, and they are not the scene. They matter because the corrections above depend on them, and because an operator looking at a number needs to know which one it is. GetTelemetry() has them all:

Field What it is on
flag_temp_c The internal shutter: the blade the camera swings in front of its own detector. "Flag" is this blade, throughout this manual
box_temp_c The camera housing
chip_temp_c The detector itself (its read-out circuit). Also called the chip temperature
fpa_temp_c The focal plane array: the detector's sensing surface

Each has a companion *_known flag, and reading it is not optional. Some cameras cannot report their internal temperatures at all (a frame format with too few metadata pixels has no room to carry them) and the field is then 0.0, which is a perfectly believable temperature. 0 °C and "no reading" are the same bytes, and only the flag tells them apart. A false chip_temp_known says nothing about the other three, which are still that frame's.

The shutter, the correction and the dark reference

An uncooled microbolometer detector drifts, and its elements drift by different amounts. Left alone, a thermal image slowly acquires a fixed pattern (some pixels reading high, some low, the whole picture sliding) that has nothing to do with the scene.

The fix is mechanical and it is the single most important thing to understand about operating one of these cameras:

The camera closes a shutter blade across its own detector, looks at that blade, and records what every pixel reads when they are all looking at the same uniform thing. That picture is the dark reference: a per-pixel offset map. Subtracting it from every subsequent frame removes the fixed pattern. The whole operation is a shutter cycle, and it is also called a non-uniformity correction (NUC). On this hardware the two are the same event: a NUC is a shutter cycle, and there is no other kind.

Three operational consequences:

A cycle interrupts the measurement for something over a second. The SDK holds the last frame taken with the shutter open and keeps delivering it, so the stream stays continuous rather than stopping. Those frames are marked is_repeated_frame, and flag_open is false on them. On a live display an operator sees a brief freeze and forgets it; in a recording it is written down and kept. Recording covers what to do about that.

A dark reference is only valid for the detector it was taken on. It describes the detector at the temperature the blade closed at. Move the detector and the reference goes stale, which is the whole of the next section.

The camera decides when to do it, and the default schedule is aggressive. The SDK runs a correction when any internal probe has drifted more than 0.1 °C, no sooner than every 12 seconds. On a camera sitting still on a bench that is a correction every twelve to sixteen seconds, for hours: measured across six cameras, 309 cycles in 184 minutes on one whose detector moved at most 0.12 K between references, and 437 in 131 minutes on one whose detector never left a 0.19 °C band. That is the probes reporting their own read-out noise rather than real drift, and optris::sdk::AutoNucConfig carries both the explanation and the two settings that fix it.

What Reading The Ambient Temperature Back Tells You

There are two ways to supply the ambient (reflected) temperature, and one flag chooses between them: let the camera work it out from its own internal probes, or measure the room yourself and say so.

// Let the camera work it out from its own internal probes.
auto radiation = imager->Radiometry().Get();
radiation.estimate_ambient = true;
imager->Radiometry().Set(radiation);

// Or measure the room yourself and say so.
radiation.estimate_ambient = false;
radiation.ambient_temp_c = 23.5f;
imager->Radiometry().Set(radiation);

The two are alternatives rather than a pair. While estimation is on, the configured value is remembered but not applied, and turning estimation off puts the stored value back into force.

What reading the ambient temperature back gives you is the configured value, in either mode. It is what was last written, never the figure the camera worked out for itself. So while estimation is on, the number you read is what was asked for and not what the measurement is using, and the two are normally different. Do not display it as "the ambient temperature in force": the SDK does not report the estimate it is using anywhere, so there is nothing to compare it against.

That is easy to miss because the value that comes back is usually plausible. A camera whose ambient has never been set reads back whatever its configuration holds, which may be 0.0, and a 0.0 meaning "never set" is indistinguishable from a 0.0 meaning "zero degrees" to anything reading the property.

Warm-up, drift and settling

A camera that has just been switched on is not yet measuring, and it does not look like it. The frames arrive at the right size with the shutter reported open and plausible numbers in them. They are wrong in two distinct ways.

Before the first dark reference exists at all, the first frame or two are converted with no offset map, so the energies run off the top of the calibration curve and clamp there. Measured in a room at about 25 °C: a PI 4x0 read a flat 125.0 °C, a PI 640i 104.3 to 125.0 °C, an Xi 400 96.9 to 125.0 °C and an Xi 410 over Ethernet 66.7 to 125.0 °C. Obviously wrong, at least.

After that the frames look right and are not. The detector is still warming, and each dark reference describes it at the temperature it was at. On a PI 4x0 from cold, with the scene held perfectly still, the reported scene mean stepped by −5.9, −7.5, −5.5, −4.2, −3.4 and −2.9 K at successive shutter cycles while the detector climbed from 42 °C to 71 °C. Each step is how wrong the frames just before it had been. An Xi 410 that was already warm held its mean to 0.6 K over the same kind of run, which is why no range check, and no comparison between frames, can stand in for simply being told.

How long does it take? As long as that camera needs, and not a fixed time. A camera that has been running holds still from the start and is measuring inside its startup calibration, about 18 seconds; the already-warm Xi 410 above reached its first reliable frame at 17.7 s. A camera switched on cold waits for its own warm-up: the PI 4x0 above took about four and a half minutes.

A PI 400 or PI 450 on USB takes 22 to 60 seconds, and may stop once more in its first minute. Such a camera is powered as long as it is plugged in, so "cold" here means unused for a quarter of an hour. Measured on a PI 450 and a PI 400 with the published SDK: when the SDK opens one, its detector reading falls by 0.2 to 1.5 K over the first half minute, and the first measurement comes when that has slowed below 0.6 K a minute, after 21.5 to 35 s on a camera used a minute earlier and 22 to 60 s after a quarter of an hour unused. In about one open in ten the fall speeds up once more right after that, and the frames of the next shutter interval, 6 to 8 seconds, are marked as not being measurements again before the camera measures for good. A flat wall in view did not move by more than 0.2 K through any of it. So expect thermal_data_reliable to turn false once more in the first minute, and do not treat that as a fault. Later in a session the check looks at two shutter intervals before it stops a camera that was measuring, so the small unevenness of the detector reading on these cameras no longer stops one for a whole interval.

"Settled" has two meanings. What the camera can decide, frame by frame, is whether the violent part of the warm-up has stopped: the 1.2 K/min a PI 450 shows in its second minute, and the 2 to 20 K/min five more cameras showed about 24 seconds after power-on. That is what the SDK reports. A full thermal settle is Optris's figure of 60 to 90 minutes, and that is an operating policy, not something the camera measures or any per-frame flag can gate. The slow tail after the first few minutes is small: the point at which a stale reference stops being distinguishable from the camera's own accuracy is about 1.8 K, inside the ±2 K a PI 4x0 specifies for itself.

So: if you need a tenth of a kelvin, let the camera warm for an hour and say so in your own interface. If you need a reading you can act on, wait for the flag in the next section.

When a frame is not a measurement

Every frame carries the answer, and the SDK withholds nothing; it tells you instead. Frames, statistics, regions and the picture all keep arriving throughout. A caller that cannot wait can show the picture, which is a perfectly good picture, and simply not put a number next to it.

There are two fields and they answer different questions.

thermal_data_reliable: "is this frame a measurement?" Check it before reporting a temperature. It is false until the startup calibration has finished and the detector has stopped warming, and false again for the whole of the recalibration that a range or mode change starts. It is one bool over several unrelated conditions, which is what makes it the right thing to gate on and the wrong thing to explain a reading with.

It is also false when the camera's raw data is not a picture at all. A detector that has stopped reading out sends the same value for every pixel. Measured on an Xi 80: every pixel of every frame read 16383, the 14-bit converter's full scale, with the shutter open and with it shut, and the dark reference built from the shut frames was 16383 everywhere too. The frame counter, the shutter and the housing temperature kept moving, so the stream looked alive, and the temperatures computed from it came out as a plausible 26 to 37 °C picture, identical on every frame. The SDK now marks every frame whose raw data is one value across the detector, and every frame converted with a dark reference that is, and logs an error naming the condition once when it starts and once when it clears. It keeps the flag false for a few frames after the data varies again, so a camera that flickers between the two states is one episode, not many.

No real scene can trigger this. Every working detector pixel has an offset of its own, so a raw frame differs from pixel to pixel by thousands of counts even when the camera looks at its own closed shutter, the most uniform scene it ever sees: across 501 frames of a PI 640i, every frame held at least 3104 different values and no single value covered more than 0.11 % of the pixels. The SDK calls a frame uniform only when all but one pixel in a thousand share one value. The check costs a few dozen memory reads on a normal frame, at any resolution.

It is also false on every frame of a camera whose calibration is missing. Without its calibration files a camera still opens and streams, and every pixel reads -100.0 °C. GetCalibrationStatus() reports kMissing in that state and GetCalibrationDiagnosis() says which files and why (see Calibration File Acquisition). The flag stays false until a calibration is loaded; for files copied in afterwards, that is the next open. SDK 12.0.22 and earlier marked those frames as measurements once the startup calibration had finished.

To see how your program handles these frames without a camera, set OTC_SIM_UNRELIABLE_FRAMES to a number of frames before starting a simulated camera. Its first that many frames after every start then read a flat 125.0 °C and carry thermal_data_reliable false, which is how a real camera's first frames look. OTC_SIM_SATURATED_FRAMES does the same for a stuck detector: the first that many frames carry raw data at 16383 on every pixel, beside a picture that still shows the scene, and are marked and logged exactly as a real camera's would be. Without either the simulated camera has no calibration to wait for, and every frame it delivers is a measurement.

detector_settling: "did the camera manage to judge its detector at all?" Three states, because two were not enough:

  • kSettled: the detector was compared across two dark references and has stopped moving.
  • kSettling: it was compared, and it is still moving.
  • kUnknown: the question could not be answered. Some frame formats carry no detector reading, and a detector count whose conversion lands outside the band a detector can physically be at is not a reading either. On such a camera the SDK deliberately withholds nothing; kUnknown is a statement about the SDK's knowledge, not a degree of unsettledness.

A third field completes the set: a region's statistics carry valid, and when it is false every temperature in them is a quiet NaN, never zero, so a reading that was never taken cannot be plotted, averaged or compared against a limit by accident. pixel_count and coverage stay truthful either way, so you can see how short the sample was. Regions has the detail.

Temperature ranges, skim and operation modes

A thermal camera does not have one measuring range. It has several, each with its own calibration, and choosing one is choosing a calibration, which is why it is not a display setting.

An operation mode is a valid combination of optic, temperature range and video format that a particular camera supports. GetDeviceInfo().operation_modes lists them. The SDK's TemperatureRange default is −20..100 °C, which is the first mode of the cameras Optris tests with and not a range every camera has; ask the device.

Skim is the detector's electrical operating point: an automatic bias trim the camera applies so the detector sits in its usable region for the range being measured. It matters here for one reason: moving the skim shifts every pixel's raw reading, so the dark reference belongs to the skim it was captured under. Change the range and both have to be re-established.

What a range change costs, measured. On an Xi 410, a −20..100 → 0..250 °C change marked 556 of the 557 frames after the write as not measurements, over 29.9 seconds; switching the extended range took 13.0 seconds, the shorter figure being the abbreviated settling a change gets when it did not move the detector's working point. The camera re-enters exactly the startup calibration a connect runs. Wait for the first frame that reports thermal_data_reliable; that flag is the whole of the signal, and there is deliberately no timer to run and nothing else to read.

Where it lands afterwards. Readings settle where opening the camera in that range lands them: about 0.7 K, re-measured on a second camera at 0.16 to 0.51 K, which is inside the accuracy the camera claims for itself. If you need a measurement immediately rather than half a minute later, open the camera in the range you want instead of changing into it.

Two refinements sit on top of a range, and both are easy to misread:

  • Extended range is for aiming, not for measuring. The extended part lies outside the calibration and reads with poor precision. It exists so that a camera with a high lower limit can be pointed and visually checked. A 150..900 °C range typically extends down to 20 °C, which is what lets a room-temperature scene be seen while in it. On a −20..100 °C range it is worth close to nothing, because the table already carries a 25 K accommodation band at each end.
  • High precision selects a finer calibration table where the mode has one. It does not change the 0.1 °C step the SDK reports, see How a temperature is stored for why. Operation Modes has what the finer table is actually worth, statistic by statistic.

Pixels that hold no measurement

Some pixels in a thermal frame are not cold and are not wrong; they hold no reading at all, and the difference matters because anything that averages, ranks or compares pixel values has to leave them out.

Unmeasurable pixels carry a reserved sentinel: kInvalidPixel in the stored format, which converts to −3376.8 °C as a float. That is far below absolute zero and is deliberately unmistakable. One of them dragged into a mean takes the whole statistic with it.

They are not an error condition. Radial distortion correction in its wide mode produces them wherever an output pixel's source falls outside the sensor: 17 380 of 110 016 pixels on a G7 optic, 15.8 % of the field, by design and on every frame.

Dead pixels are a different thing: detector elements with no usable calibration. The camera and the SDK know where they are, and the pipeline fills them from their neighbours, searching outward in a fixed pattern, skipping the outer border and other dead pixels, and taking the mean of four substitutes' corrected values. A camera may also accumulate dynamic dead pixels over its life, which the SDK keeps in a file of its own per serial; see Data Files.

The safe way to read a whole frame as temperatures is ThermalFrame::CopyTemperaturesTo(), which answers NaN for a pixel it could not measure. Applying the conversion formula yourself turns that into a plausible-looking −3376.8 °C instead. Reading temperatures out of a frame has the detail.

Two corrections that are about the optics

Both are on by default, and both are about the lens rather than the detector.

Size-of-source correction deals with a real optical effect: a hot target does not land only on the pixels it geometrically covers. Some of its flux scatters inside the optic and arrives as a faint halo over the surrounding pixels, so a small hot object reads too cool and its surroundings read too warm. The correction subtracts that halo back out.

It does nothing until you give it a strength, and that is honest rather than an oversight. The correction table a calibration ships is a shape (where the stray flux comes from) normalised so that it says nothing about how much of it there is. Two optics with the same halo geometry and different scatter ship the same table, so the magnitude cannot be recovered from it by any arithmetic. It has to be measured for a camera and its optic and set as sos_strength; until it is, the correction is an exact identity. Processing().Correction() says so explicitly rather than reporting a correction that is on and quietly doing nothing.

Radial distortion correction straightens the picture. A wide lens bends straight lines, more the further from the centre; the correction resamples the frame to a rectilinear one. Three modes: off, which keeps the lens's own geometry; normal, the default; and wide, a table tuned for wide-angle optics, which is the mode that leaves the unmeasurable border described above. Where a calibration carries no coefficients it is a no-op rather than an error.

Frame Processing Settings covers both, along with the noise filters, which, unlike detail enhancement, run inside the radiometric path and do change reported temperatures.

Showing a temperature in Fahrenheit

The SDK measures and stores in degrees Celsius, and every temperature difference in kelvin. Frames, regions, alarm limits, PIF mappings, the colour scale, radiation parameters and recordings all stay that way, so a stored limit or a recording never depends on the unit somebody was looking at. The unit is for the moment a number is shown to a person or typed by one, and the SDK does that part too, so every Optris application and every SDK program converts, rounds and labels a temperature the same way.

A temperature and a difference are converted differently. A temperature is a point on the scale: °F = °C × 9/5 + 32. A difference is a distance on it: °F = K × 9/5, with no offset. An alarm's hysteresis, the limits of a rule with a reference (which are offsets), a standard deviation and a colour-scale span are differences. Converted like a temperature, a hysteresis of 2 K reads 35.6 °F; it is 3.6 °F. So there are two pairs of functions, and the name says which kind of number each takes.

Converting never rounds; showing rounds once. A value a person types in Fahrenheit is converted to Celsius and stored as it is: 70.1 °F is 21.1666… °C, and it shows as 70.1 °F again. Rounding the Celsius value to a tenth first would store 21.2 °C, which reads back as 70.2 °F. Format converts and then rounds once, in the unit shown, half away from zero: 21.25 shows as 21.3 and -21.25 as -21.3 at one decimal. The half is judged on the number as it is written: 2.675 rounds to 2.68 at two decimals, although a computer stores 2.675 as a value a hair below it (most languages' own rounding gives 2.67 there). A result that rounds to zero is shown without a minus sign. A missing reading (NaN) is an empty string, so your application decides how to show it.

using optris::sdk::Temperatures;
using optris::sdk::TemperatureUnit;
const auto unit = TemperatureUnit::kFahrenheit;

// A reading, with the symbol: "70.3 °F".
const std::string shown =
    Temperatures::Format(21.3, unit, 1) + " " + Temperatures::Symbol(unit);
// A hysteresis of 2 K: "3.6", not "35.6".
const std::string band = Temperatures::FormatDifference(2.0, unit, 1);
// What the person typed, stored in Celsius without rounding.
const double limit_c = Temperatures::ToCelsius(70.1, unit);

In C#, Format takes an IFormatProvider as its last argument: a German culture writes 21,3, and leaving it out uses the current culture, as everywhere in .NET. C, C++, Python and Rust take the decimal separator, .'or','. There is never a thousands separator, so 3000 °C is5432.00in Fahrenheit. A unit other than Celsius and Fahrenheit, or more than 15 decimals, is refused with the SDK's invalid-argument error. Parsing what a person typed stays with your application: read the number your way, then callToCelsius()for a temperature orToKelvinDifference()` for a difference.

A CSV snapshot can be written in Fahrenheit too: see A CSV in degrees Fahrenheit.

The text the SDK draws itself

Some temperatures are turned into text by the SDK rather than by your application: the reading in a region's label drawn into the picture (show_label in the region's style), and the sentence a JPEG snapshot opens with ("The scene measured 21.3 C to 36.6 C."). One setting on the camera chooses the unit for all of them. It is display only: everything the API takes and returns stays in degrees Celsius, and so does everything written for a program to read (the radiometric PNG and TIFF, the metadata in every snapshot, the log). A CSV snapshot keeps its own unit, set when it is written.

using optris::sdk::TemperatureUnit;

// Region labels read "INLET 70.3F" instead of "INLET 21.3C", and a
// JPEG snapshot says what the scene measured in Fahrenheit.
imager->SetDisplayTemperatureUnit(TemperatureUnit::kFahrenheit);

// The numbers the API hands back are still Celsius.
const float max_c = frame.GetMaxTemperature();

// Celsius, the default, draws exactly what it always drew.
imager->SetDisplayTemperatureUnit(TemperatureUnit::kCelsius);

The reading in a label is converted and rounded by Format, at one decimal, so a label and your own text for the same reading always agree, and no locale changes it. The label carries the letter C or F without a degree sign, as before. The setting works the same on a recording being replayed and on a still opened from a snapshot (a still draws no region labels, so there it changes the JPEG sentence alone), and it takes effect on the next frame drawn. With several cameras, each has its own setting: a canvas is drawn by the camera it is set on, so its labels follow that camera.

Text the SDK writes for a program or a developer is not affected: log lines, error messages and the configuration files all stay in degrees Celsius, because they are read back by something that expects it.

The ready-made viewers have the same switch. simpleView, showcase and the .NET SimpleView take --fahrenheit on the command line and switch with the U key (showcase also has a "Units" row in its menu); the playground sets it in src/playground.cc, experiment 2b, and switches with U too. Of the command line tools, otc_stream and otc_snapshot take --fahrenheit for their CSV and JSON output; see the command line tools' conventions.

How a temperature is stored

The SDK carries every thermal pixel as a 16-bit integer:

internal = (celsius * 10) + 1000
celsius  = (internal - 1000) / 10

That is where the 0.1 °C step in every reported temperature comes from. It is a property of the representation, not of the detector, which is why selecting a finer calibration table does not make the reported numbers finer.

The format's ceiling is 3176.7 °C; a conversion above it saturates there rather than wrapping. There is no corresponding cold saturation, because the sentinel for "no measurement" sits more than 31 000 counts below the coldest temperature anything can be at. CelsiusToInternal() refuses rather than wrapping, and IsEncodableTemperature() says which of the three outcomes you got.

Glossary

Term Meaning
ADU Analogue-to-digital unit. The detector's raw integer reading, before any calibration
Ambient temperature The temperature of the surroundings. In this SDK it is also the reflected reference; there is one setting for both
Characteristic curve The per-unit, per-optic, per-range calibration that turns energy into a temperature
Chip temperature The detector's own temperature. Also called the detector temperature
Dark reference The per-pixel offset map captured while the shutter blade covers the detector. Valid only for the detector temperature and skim it was taken at
DDE Detail enhancement. A display-only sharpening; it does not change a reported temperature
Emissivity (ε) How efficiently a surface radiates, 0 to 1. The single most important measurement setting
Energy The linear radiometric quantity between raw counts and temperature
Flag The camera's internal shutter blade. flag_open is whether it is out of the way; flag_temp_c is its temperature
FPA Focal plane array: the detector's sensing surface
NUC Non-uniformity correction. On this hardware, one shutter cycle; the thing that captures a dark reference
Operation mode A valid combination of optic, temperature range and video format for a given camera
RDC Radial distortion correction. Resampling the frame to a rectilinear picture
Skim The detector's electrical operating point, trimmed automatically per temperature range
Size-of-source (SoS) The halo of stray flux a hot target spreads over its surroundings inside the optic, and the correction that subtracts it
Transmissivity (τ) How much of the object's radiation survives the path to the camera, 0 to 1

Next: Features for what the SDK does with all of this, and Using the API for how to drive it.