OptrisOptrisDocumentation

Documentation/OTC SDK/optris_sdk — Python API referencev12.0

OperationMode

class optris_sdk.control.OperationMode

One combination a camera can be set to, whole.

Optics, temperature range, resolution and frame rate are not independent knobs: a camera offers a list of combinations and you pick one. That is why this is a record read from Camera.operation_modes rather than four setters.

temp_min_c, temp_max_c and extended_range_c are in the SDK's temperature unit (Celsius unless optris_sdk.set_temperature_unit chose Fahrenheit): a -20..100 C mode reads -4..212 in Fahrenheit, and passes back to Camera.operation_mode as it was read.

OperationMode — properties

optics_fov

optics_unit

What optics_fov is measured in.

One of OPTICS_UNIT_DEGREES (a field of view), OPTICS_UNIT_MILLIMETRES (a focal length) or OPTICS_UNIT_UNKNOWN. Read it before printing a unit: the number alone cannot tell a 29 mm lens from a 29-degree one. optics_unit_symbol turns it into text.

optics_text

temp_min_c

temp_max_c

width

height

framerate

extendable

Whether the range extension is available in this mode.

high_precision_available

Whether this range's calibration offers high precision.

Nothing to do with frame rate; see Camera.high_precision_enabled.

extended_range_c

The span the extension reaches, as (min, max).

OperationMode — methods

__repr__()