ReferenceCorrection
class optris_sdk.control.ReferenceCorrection
The reference-body correction in force.
- Attribute
active— whether one is applied. False both for a source with no correction set and for one that cannot carry one --Detector.reference_correction_supportis what separates those two. - Attribute
reference_c— what the reference body was said to be, in the SDK's temperature unit (Celsius unlessoptris_sdk.set_temperature_unitchose Fahrenheit). - Attribute
measured_c— what the camera read there, in the same unit. - Attribute
gain_fit— whether a gain fit was applied rather than an offset alone. Chosen by the SDK when the table was rebuilt, from how far the reference sat from the shutter temperature, so the setter cannot report it and this is where it is read.