DetectorSettling

public enum DetectorSettling

What the SDK concluded about whether this camera's detector has warmed up and stopped drifting.

Three states, because two were not enough. The SDK waits for a detector to stop moving before it calls a frame a measurement: a dark reference describes the detector at the temperature it was taken at, so one taken while the detector was still warming makes every frame after it wrong by however far the detector has since moved. Measured on a PI 4x0 (serial 16030066) from cold, the reported scene mean stepped by -5.9, -7.5, -5.5, -4.2, -3.4 and -2.9 C at successive shutter cycles while the detector climbed from 42 C to 71 C.

This is not "fully settled". Optris's own figure for a full thermal settle is 60 to 90 minutes, and that is a policy statement rather than something the SDK measures. What the criterion answers is narrower and is the question a per-frame flag can answer: has the violent part of the warm-up stopped. If your application needs the hour, it owns that clock.

Value
Unknown 0 The SDK cannot judge this camera's detector, because no detector reading reaches the criterion.
Settling 1 The criterion applies and the detector has not stopped moving.
Settled 2 Two consecutive dark references were taken at the same detector temperature, at a rate that says it has stopped.