DetectorSettling

pub enum DetectorSettling

What the SDK concluded about whether this camera's detector has warmed up and stopped drifting.

Three states, because two were not enough. A dark reference describes the detector at the temperature it was taken at, so one taken while the detector was still warming makes every frame after it wrong by however far the detector has since moved, measured on a PI 4x0 (serial 16030066) from cold, the reported scene mean stepped by -5.9, -7.5, -5.5, -4.2, -3.4 and -2.9 °C at successive shutter cycles while the detector climbed from 42 °C to 71 °C.

Settled is not "fully settled". Optris's own figure for a full thermal settle is 60 to 90 minutes, which is policy rather than something this SDK measures; what the criterion answers is whether the violent part of the warm-up has stopped. An application that needs the hour owns that clock.

DetectorSettling — variants

Unknown

The SDK cannot judge this camera's detector, because no detector reading reaches the criterion.

Not a degree of unsettledness: the question was not answerable, rather than answered badly. A frame format with too few metadata pixels carries no detector reading at all (a PI 160x120 at 100 Hz has two), and a count whose conversion does not finish inside the physical band is not a reading either. Nothing is withheld in this state, and it is also what a player and the simulated camera report.

Settling

The criterion applies and the detector has not stopped moving. The only state that makes thermal_data_reliable false on its own account.

Settled

Two consecutive dark references were taken at the same detector temperature, at a rate that says it has stopped.