public enum SignalField

Which quantity a process-interface channel or an alarm rule refers to.

Which values are meaningful depends on the SignalDomain they are paired with: a region has a minimum and a mean, the device has telemetry, a process-interface channel has a value. Pairing a domain with a field from another group is rejected when the rule is added.

A process-interface channel accepts only the region statistics.

A field's value is in the unit of the quantity it carries: the temperatures (a region's minimum, maximum, mean, median and percentiles, the camera's probe, ambient and scene temperatures) in the SDK's temperature unit (Celsius unless Sdk.TemperatureUnit chose Fahrenheit), the standard deviation as a difference in degrees of it. Counts, areas, coordinates, coverages and states are not converted.

Value
Min 0 The coolest pixel.
Max 1 The warmest pixel.
Mean 2 The mean over the region.
Median 3 The median. Requires percentiles to be enabled.
StdDev 4 The standard deviation.
Coverage 5 How much of the region had sensor data behind it.
Area 6 The region's area.
HotSpotX 7 Where the warmest pixel is, horizontally.
HotSpotY 8 Where the warmest pixel is, vertically.
ColdSpotX 9 Where the coolest pixel is, horizontally.
ColdSpotY 10 Where the coolest pixel is, vertically.
BlobCount 11 On a RegionKind.BlobGroup, how many blobs were detected this frame. On every other kind, how many children the region has -- all of them, not only blobs.
ThresholdShare 12 The fraction of the region's measurable pixels past its threshold, 0 to 1 -- not a percentage. Published only while the region has Region.Threshold enabled; a rule on a region that has not reports AlarmState.Orphaned rather than alarming on a share of zero that means "not counted".
P01 13 The region's 1st percentile. Published only while the region has Region.ComputePercentiles set and produced a measurement; a rule on it otherwise reports AlarmState.Orphaned rather than alarming on a zero that means "not computed". The same rule SignalField.Median follows, since both come from the one selection.
P05 14 The region's 5th percentile. See SignalField.P01.
P95 15 The region's 95th percentile. See SignalField.P01.
P99 16 The region's 99th percentile. The robust hot value: a few noisy or dead pixels do not trip it the way they trip SignalField.Max, and it is not the same number. See SignalField.P01 for when it is published.
FlagTemperature 100 The shutter flag's temperature, published only while the frame carries a shutter reading.
BoxTemperature 101 The camera housing's temperature, published only while the frame carries a housing reading -- the same rule SignalField.FlagTemperature follows, from the same decoder. See FrameMetadata.HousingTemperature.
ChipTemperature 102 The detector's temperature, published only while the frame carries a detector reading.
FpaTemperature 103 The focal-plane array's temperature. No producer: nothing publishes this, so a rule on it can never fire.
AmbientTemperature 104 The ambient temperature the camera is using.
SceneMinimum 105 The coolest pixel in the whole frame.
SceneMaximum 106 The warmest pixel in the whole frame.
FocusPosition 107 Where the lens is focused. No producer: nothing publishes this, so a rule on it can never fire.
FrameRate 108 Frames per second. No producer: nothing publishes this, so a rule on it can never fire.
DroppedFrames 109 How many frames the SDK has dropped. No producer: nothing publishes this, so a rule on it can never fire; use SignalField.DeliveryShare instead.
FlagOpen 110 Whether the shutter is open. Non-zero for open.
NucInProgress 111 Whether a shutter cycle, and so the correction inside it, is running. Non-zero while it is.
Connected 112 Whether the camera is connected.
ProbeTemperature 113 The external temperature probe's reading, in the SDK's temperature unit.
DeliveryShare 114 How much of what the camera took is reaching the pipeline: 0 to 1, where 1 is every frame.
DigitalIn 200 A digital input's level.
DigitalOut 201 A digital output's level.
AnalogIn 202 An analog input's decoded value, in its channel's own unit: a temperature in the SDK's temperature unit on a temperature channel.
AnalogOut 203 The level an analog output has on its wire, in volts or milliamperes: the same number in Celsius and in Fahrenheit, even on a temperature channel.
ProcessInterfaceFlag 204 The process interface's flag line.
FailSafe 205 The fail-safe watchdog's state.
MotionBlobCount 300 How many moving blobs a zone sees.
MotionArea 301 Their combined area.
MotionVelocity 302 The fastest of them.
Value 400 An application-published value.