RadiationConfig
class optris_sdk.control.RadiationConfig
Everything that turns radiance into a temperature.
Returned by Radiometry.get and accepted by
Radiometry.set. A whole record on purpose: the parameters
are applied together, and reading one before changing another is
what stops a partial write turning into a default.
- Attribute
emissivity— how much of a black body's radiation the target emits, 0..1. - Attribute
transmissivity— how much of it survives the path to the lens, 0..1. 1.0 unless something is in the way. - Attribute
ambient_temperature_c— the reflected reference, in the SDK's temperature unit (Celsius unlessoptris_sdk.set_temperature_unitchose Fahrenheit). The camera holds one temperature for both ambient and reflected. A fresh record's default follows the unit too: 20 C reads 68 F. - Attribute
estimate_ambient— let the camera estimate the ambient rather than believing the number above. - Attribute
emissivity_affects_image— whether a region's own emissivity changes the picture as well as the numbers. On by default. The image half runs on both pipelines, GPU and CPU, and so on a simulated camera too; on playback and a still, which have no pipeline, a region reportsEMISSIVITY_READING_ONLYhowever this is set. - Attribute
emissivity_own_scale— whether each corrected region is coloured on its own temperature range instead of the scene's. Off by default, and only meaningful whileemissivity_affects_imageis set. Works on every source: the rescaling is a pass over the delivered picture, not a pipeline stage. This is what producesEMISSIVITY_OWN_SCALE.