PifFailSafeConfig

pub struct PifFailSafeConfig

The fail-safe line: a watchdog that must keep changing, plus the conditions that trip it.

The toggle is the load-bearing part. The line alternates every period_ms; a receiver that sees it stop knows the host stopped, which is the entire point. A constant level would be indistinguishable from the fault it exists to signal.

PifFailSafeConfig — fields

enabled: bool

Arm the watchdog.

on_alarm: bool

Never implemented, on any camera. Setting this is refused with Status::NotSupported rather than accepted and ignored.

The field exists so that a caller porting code that asked for it is told so, instead of finding the option absent and assuming the fail-safe line covers alarms. Drive an ordinary digital or analog output from the alarm rule instead, and keep the fail-safe line for camera and pipeline health.

on_flag_timeout: bool

Not configured from here, and setting this is refused with Status::NotSupported.

The shutter is watched by the camera, not by the process interface, through a legacy setting reachable only from a legacy configuration file -- not from Rust, from .NET or from the C ABI. Refused rather than accepted so that no application believes it has a shutter interlock it does not have.

index: u8

Which channel carries the line.

signal: PifSignal

The line's electrical nature.

active_value: f32

Level driven when the watchdog is satisfied.

inactive_value: f32

Level driven when it is not.

period_ms: u32

Watchdog toggle period, in milliseconds.

flag_cycles: u16

Inert. Meaningful only alongside PifFailSafeConfig::on_flag_timeout, which is refused, so this can never be reached.

Deliberately not refused on its own: it is part of the record every caller round-trips, and refusing a value that was merely read back and written again would reject a plain watchdog that asked for nothing.

default_to_range_endpoints: bool

Beat between the two ends of the channel's range when no level was given. True by default.

With this set and active_value / inactive_value still at their defaults of 1 and 0, an analog fail-safe drives the top and the bottom of the channel's output range -- 20 and 4 mA, 20 and 0 mA, 10 and 0 V -- the largest edge the channel can make. Set either level to anything else and that is what is driven instead: this governs the default, it never overrides an instruction.

Clear it to drive a literal 1 and 0 on an analog line, which on a live-zero range is then refused as a flat line. Inert on a digital line and on the dedicated fail-safe terminals.

PifFailSafeConfig — methods

pub fn new() -> Result<Self>

The SDK's defaults: disabled, on channel 0, a 500 ms toggle period, tripping on neither alarms nor flag timeout.

Errors

Returns an error only if the SDK's own initialiser fails.