SignalId
public readonly record struct SignalId : IEquatable<SignalId>
Which observation an alarm rule watches.
A rule is one signal, one condition, and what to do about it, so a rule on a measuring region's temperature, on a process-interface input and on a value your own application publishes are the same shape. This is the part that says which.
Use the factory methods rather than setting the fields by hand: they pair a domain with a field from the right group, which is the mistake this type otherwise invites.
SignalId: properties
public SignalDomain Domain { get; init; }
Where the observation comes from.
public SignalField Field { get; init; }
Which quantity of it.
public uint Index { get; init; }
Which one: a process-interface channel number, a motion zone id, or the slot from ThermalCamera.RegisterSignal(…).
Left at zero for a region signal: the camera fills it in from the region the rule is attached to, both when the rule is added and again whenever it is updated, so a caller never has to look one up.
Zero is never a live region. To name a region other than the one the rule is attached to (an inhibit watching a different area, say), pass its slot from ThermalCamera.GetRegionSignalSlot(…) to SignalId.Region(…).
SignalId: methods
public static SignalId Device(SignalField field)
A camera telemetry reading.
field: Which reading.
Returns: The signal.
public static SignalId Motion(uint zoneId, SignalField field)
What a motion zone saw.
zoneId: The zone.field:SignalField.MotionBlobCount,SignalField.MotionAreaorSignalField.MotionVelocity.
Returns: The signal.
public static SignalId ProcessInterface(uint channel, SignalField field)
A process-interface channel.
channel: Its index.field: Which quantity:SignalField.DigitalIn,SignalField.AnalogInand so on.
Returns: The signal.
public static SignalId Region(SignalField field)
A statistic of the region the rule is attached to.
field: Which statistic.
Returns: The signal.
The camera resolves the region when the rule is added and again on every update, so this stays bound to its own region across a threshold edit.
public static SignalId Region(uint slot, SignalField field)
A statistic of one particular region.
slot: Its packed signal slot, fromThermalCamera.GetRegionSignalSlot(…).field: Which statistic.
Returns: The signal.
Use this for an inhibit or a reference that names a region other than the rule's own. The slot is a runtime handle: it changes if the region is removed and re-added, so resolve it from the region id at the point of use rather than storing it. A rule whose slot has gone stale reports AlarmState.Orphaned rather than measuring whichever region took the slot over.
public static SignalId User(uint slot)
A value the application publishes.
slot: The slot fromThermalCamera.RegisterSignal(…).
Returns: The signal.